- Radiology Equipments & Accessories[10]
- Remote Control[7]
- Ultrasonic[10]
- Clinical Analytical Instruments[10]
- Emergency & Clinics Apparatus[10]
- Other Electronic Components[4]
- Physical Therapy Equipments[4]
- Spectrometers[7]
- Laser Equipment[2]
- Other Lights & Lighting Products[1]
- General Assay & Diagnostic Apparatus[10]
- Dental Equipments[10]
- Surgical Instruments[10]
- Separation Equipment[8]
- Machinery Design Services[3]
- Hospital Beds[4]
- Sterilization Equipments[6]
- Drying Equipment[3]
- Water Filter[2]
- Medical Consumables[10]
- Ward Nursing Equipments[4]
- Bidets[1]
- Weighing Scales[1]
- Healthcare Supplement[1]
- Vacuum Cavitation System[2]
- Contact Person : Ms. Yi Nancy
- Company Name : Guangzhou Maya Equipment Co., Ltd.
- Tel : 0086-20-39160381
- Fax : 0086-20-39154381
- Address : Guangdong,Guangzhou,Room 409 ,Daxin business plaza, Shangjiao, Dashi town ,Panyu area, Guangzhou city China
- Country/Region : China
MA2000B2 ventilation machine
Ventilator:
1. Display mode High-definition 8.4” TFT color LCD screen display
2. Ventilation mode IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH
3. Ventilation parameters Tidal volume 50~1500 ml
Rate 2~99 bpm
SIMV rate 2~20 bpm
I:E 2:1~1:8
Inspiratory trigger pressure -1.0~2.0kPa
PEEP 0~2.0 kPa
Pressure range 1.0~6.0 kPa
SIGH 1.5 times the inspiratory time
Inspiratory platform 0~50%
4. Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform
5. Monitoring of oxygen concentration 21%~100%
6. Security Alarm System
Oxygen Concentration alarm upper limit setting range 50%~100%
low limit setting range 15%~50%
Airway pressure alarm upper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa
Per-minute ventilation volume alarm upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
Sustained high-pressure alarm it will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation alarm it will give sound and light alarm if there is no tidal volume input for 15 seconds
Power alarm
6. Power AC 220V±10% 50Hz±1Hz
UPS, Storage battery
7. Mechanical Arms
8. End-expiratory CO2 monitoring (choose and match)